大鼠全身体积描记检测系统的性能确认
作者:
作者单位:

山东省药学科学院,山东省化学药物重点实验室,济南 250101

中图分类号:

R-33


Performance qualification of whole-body plethysmography in rats
Author:
Affiliation:

Shandong Academy of Pharmacetical Sciences, Shandong Provincial Key Laboratory of Chemical Drug, Jinan 250101, China

  • 摘要
  • | |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • | |
  • 文章评论
    摘要:

    目的 选用巴氯芬和氨茶碱对 SD 大鼠全身体积描记检测系统进行性能确认。 方法 将 30 只动物分为空白对照组、30 mg / kg 巴氯芬组和 30 mg / kg 氨茶碱组,并用全身体积描记系统采集给药前和给药后 1 h、2 h 和 4 h 的呼吸频率(f)、潮气量(TVb)和每分通气量(MVb)。 并重复以上实验 2 次以检验系统稳定性。 结果 空白对照组各指标无明显变化,与空白对照组比较,给予 30 mg / kg 巴氯芬后 1 h、2 h、4 h,SD 大鼠的 f 显著降低,TVb 显 著升高(P<0. 05);给予 30 mg / kg 氨茶碱后 1 h、2 h、4 h,SD 大鼠的 f 和 MVb 显著升高(P<0. 05)。 两次重复实验的结果与第 1 次结果一致。 结论 全身体积描记检测系统可以及时有效的监测到大鼠给予受试物后引起的呼吸功能指标的变化,本系统可以应用到非临床安全药理实验研究中。

    Abstract:

    Objective To research the performance qualification of whole-body plethysmography in SD rats with baclofen and aminophylline. Methods Totally 30 Sprague-Dawley rats were divided into three groups: control group, baclofen group, and aminophylline group. The baclofen and aminophylline group received a single dose of 30 mg / kg baclofen or aminophylline in the study. The tidal volume box(TVb), minute ventilation box(MVb), and respiratory rate ( f)were detected with whole-body plethysmography at before dosing and 1 h, 2 h, 4 h after dosing. And tested the stability of the system through repeated the study twice. Results Control group had no significant change on the respiratory parameters.Compared with control group, TVb of the baclofen group increased significantly at 1, 2, 4 h after dosing(P< 0. 05), And f decreased at 1, 2, 4 h after dosing ( P< 0. 05). F and MVb of the aminophylline group increased significantly at 1, 2, 4 h after dosing(P< 0. 05). The result of two repeated studies were consistent with the first data. Conclusions The whole-body plethysmography can detect relevant changes in the respiratory system in SD rats after dosing. It can be used in conducting respiratory safety pharmacological investigation.

    参考文献
    相似文献
    引证文献
引用本文

张燕华,郑智勇,李海涛,陈 凯.大鼠全身体积描记检测系统的性能确认[J].中国比较医学杂志,2021,31(2):102~106.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2020-06-24
  • 在线发布日期: 2021-04-07
防诈骗提示!请勿点击不明链接或添加个人微信。编辑部所有邮箱后缀均为@cnilas.org