Consideration of Physical Parameters of Individually Ventilated Cages
Clc Number:

R-33

  • Article
  • | |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • | |
  • Comments
    Abstract:

    Objective To formulate national or trade standards of individually ventilated cages(IVC),physical parameters of IVCs from different factories were tested.Methods\ The IVCs were tested according to the standards of GB 14925,90/219/EEC and 98/81/EC.Results\ Compared with barrier facility,the noise of the IVCs is higher;parameters of gradient pressures and cleanse were greatly floating.Conclusion\ Standards of IVC should be carefully determined and can not be simply copied from those of laboratory animal barrier facilities.

    Reference
    Related
    Cited by
Get Citation
Related Videos

Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:April 26,2006
Article QR Code